Klein Tools MM2300A Handleiding

Hier is de Klein Tools MM2300A (Multimeter) gebruikershandleiding. 10 pagina's in taal met een gewicht van 4,949,920.0 Mb. Als u geen antwoorden op uw probleem kunt vinden Vraag het onze community.

Pagina 1/10
CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
customerservice@kleintools.com
www.kleintools.com
139898 Rev. 11/14 C
139898 Rev. 11/14 C
ENGLISH
TRMS
AUTO / MANUAL
RANGE
AUTO HOLD
MAX / MIN
BAR GRAPH
BACKLIGHT
LEAD HOLDER
3-5/6 DIGIT
6000 COUNT LCD
ENGLISH
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range Resolution Accuracy
600mV to 600V
0.1mV to 0.1V
± (0.5% + 4 digits)
1000V
1V
± (0.8% + 10 digits)
Overload Protection:
1000V
Input Impedance (Nominal)
: 10MΩ, <100pF
Input Impedance (Low Z)
: 3kΩ, <200pF, up to 600V RMS
AC Voltage Measurement
Range Resolution
Accuracy
40Hz to 400Hz
Accuracy
400Hz to 5kHz
600mV
0.1mV
± (0.75% + 5 digits)
± (2.0% +3 digits)
6V to 600V
1mV to 0.1V
± (2.0% +3 digits)
1000V
1V
± (0.75% + 8 digits)
± (2.0% +3 digits)*
continued on next page.
Overload Protection:
1000V RMS
Input Impedance (Nominal):
10MΩ, <100pF
Input Impedance (Low Z)
: 3kΩ, <200pF, up to 600V RMS
Frequency Response:
40 to 5kHz
Response:
True RMS
*Accuracy for 400Hz to 1kHz range only.
DC Current Measurement
Range Resolution Accuracy
600µA to 400mA
0.1µA to 0.1mA
± (1.0% + 5 digits)
6A to 10A
1mA to 10mA
± (1.5% + 5 digits)
Overload Protection: µAmA: F440mA/1000V fuse / A Input: F11A/1000V fuse
AC Current Measurement
Range Resolution
Accuracy
40Hz to 2kHz
600µA to 400mA
0.1µA to 0.1mA
± (1.2% + 5 digits)
6A to 10A
1mA to 10mA
± (1.8% + 5 digits)
Overload Protection: µAmA: F440mA/1000V fuse / A Input: F11A/1000V fuse
Frequency:
40 to 2kHz
Response:
True RMS
Resistance Measurement
Range Resolution Accuracy
600
Ω
to 6M
Ω
0.1
Ω
to 0.001M
Ω
± (1.0% + 5 digits)
60M
Ω
0.01M
Ω
± (1.5% + 10 digits)
Overload Protection:
600V RMS
Capacitance Measurement
Range Resolution Accuracy
60nF
0.01nF
± (3.5% + 6 digits)
600nF to 6000µF
0.1nF to 1µF
Overload Protection:
600V RMS
Frequency Measurement
Range Resolution Accuracy
99.99Hz to 399.9kHz
0.01Hz to 0.1kHz
± (0.1% + 3 digits)
Overload Protection:
600V RMS
Sensitivity:
> 2.5V RMS
Duty Cycle Measurement
Range Resolution Accuracy
1.0 to 99.0%
0.1%
± (0.2% per kHz + 0.1% + 5 digits)
Overload Protection:
600V RMS
Sensitivity:
> 2.5V RMS
Frequency Range:
<10kHz
Diode Test
Overload Protection Range Test Current Open Circuit Voltage
600V RMS
6.0V
Appx. 0.25mA
<
1.8V DC (typ.)
Continuity Test
Overload Protection Open Circuit Voltage Threshold (Appx.)
600V RMS
<
0.44V
<40
Ω
Temperature Measurement
Range Resolution Accuracy
-50 to 600°C
0.1°C
± (1.0% + 3.0°C)
600 to 1000°C
1°C
± (1.0% + 3.0°C)
-58 to 600°F
0.1°F
± (1.0% + 5.4°F)
600 to 1832°F
1°F
± (1.0% + 5.4°F)
Overload Protection:
600V RMS
Thermocouple Accuracy:
Not specified
Specifications subject to change.
Do Not apply voltage to thermocouple.
-50° to 1000°C (-58° to 1832°F)
8. Temperature:
Features:
Testing Fuse:
When selecting uA or mA/A functions the meter will indicate
LEAd
” if the test leads are not properly connected to the meter.
Note: This message remaining on the screen after properly
connecting the test leads is an indication that the fuse is open.
7. Frequency (Hz) / Duty Cycle: <500kHz
Features:
HOLD
6. Low Impedance (Low Z): < 600V
DO NOT use (LOW Z) mode at voltages greater than 600V.
Press SELECT to choose AC or DC Voltage.
Use Low Impedance (LOW Z) mode to reduce ghost voltages
See AC/DC Voltage instruction
Capacitance Features:
HOLD
Safely discharge capacitor before
measurement.
Reading may take up to 60 seconds
for large capacitors.
AUTO HOLD
MAX/MIN REL
HOLD
SYMBOLS USED ON LCD
AC Measurement
DC Measurement
-
Negative DC Value
AT
Auto Range Active
O
.L.
Overload: Range Exceeded
Apo
Auto Power-Off Active
Low Battery
HOLD
Hold Active
MIN
Minimum Reading
M
AX
Maximum Reading
%
Duty Cycle Mode Frequency Mode
V
Voltage Measurement
A
Current in Amps
Ω
Resistance in Ohms Diode Test
F
Capacitance in Farads Continuity Test
Relative / Zero Mode
n
Nano 10
-9
µ
Micro 10
-6
m
Milli 10
-3
k
Kilo 10
3
M
Mega 10
6
°F
Degrees Fahrenheit
°C
Degrees Celsius
A-HOLD
Auto-Hold Active
LOW
Z
Low Impedance
WARRANTY
www.kleintools.com/warranty
CLEANING
Clean the instrument by using a damp cloth. Do not use abrasive cleaners
or solvents.
Take care to keep the sensor lens clean at all times. Clean the lens using a soft
cloth or cotton swab with water or rubbing alcohol only, and allow the lens to
dry before use.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
DISPOSAL / RECYCLE
Do not place equipment and its accessories in the trash. Items
must be properly disposed of in accordance with local regulations.
1000V
10A
PRESS
PRESS
°F
°C
MM2300A
INSTRUCTION MANUAL
Electrician’s / HVAC
TRMS Multimeter

Probleemoplossing Klein Tools MM2300A

Als je de handleiding al zorgvuldig hebt gelezen maar geen oplossing voor je probleem hebt gevonden, vraag dan andere gebruikers om hulp


Specificaties

Merk: Klein Tools
Categorie: Multimeter
Model: MM2300A